FIGURE

Figure 8

ID
ZDB-FIG-221020-103
Publication
Pérez-Davila et al., 2022 - 3D-Printed PLA Medical Devices: Physicochemical Changes and Biological Response after Sterilisation Treatments
Other Figures
All Figure Page
Back to All Figure Page
Figure 8

Cell viability detected in mouse fibroblast cells (NCTC clone 929) in the presence of different extracts of PLAGR, PLASCCO, PLALTSF, PLAHPGP and PLASS discs, compared to the positive control (phenol) and negative control (DMEM). Results are expressed as percentage compared to the negative control ± standard error. Statistical significance was determined to * (p ≤ 0.05).

Expression Data

Expression Detail
Antibody Labeling
Phenotype Data

Phenotype Detail
Acknowledgments
This image is the copyrighted work of the attributed author or publisher, and ZFIN has permission only to display this image to its users. Additional permissions should be obtained from the applicable author or publisher of the image. Full text @ Polymers (Basel)